Variable Energy X-ray Fluorescence Source
Journal of Instrumentation(2023)
摘要
We detail the design of a variable energy, x-ray fluorescence source using a low activity (1.8x10(6) dpm) Tc-99 beta source that irradiates thin foils. By rotating the source among foils of Ti, Zn, Nb, Ag, and Au, the device produces x rays between 4 and 70 keV at a rate near 1 Hz. When the source is placed in a storage position, the external radiation is non-detectable. The design of the shielding and rotation mechanism permits use in vacuum and at liquid nitrogen temperature. The design is intended for the study of the low energy response to radiation impinging upon Ge detector surfaces. The source will be useful for understanding the detector response in large-scale Ge arrays such as MAJORANA and LEGEND.
更多查看译文
关键词
X-ray fluorescence (XRF) systems,Solid state detectors,Cryogenic detectors
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要